Gold Film Mercury Vapor Analyzer
The Jerome® J405 portable mercury vapor monitor is redefining the portable mercury vapor analyzer market. Significant performance enhancements and modern communications capabilities have been combined to create a new option for mercury spill detection and clean-up analysis. The J405 utilizes the industry-proven, inherently stable and reliable gold film sensor technology and simple, one-button operation. The ergonomically-designed handle, lighter exterior case and significantly lower detection capabilities (0.5 µg/m3) make it a portable, easy to use, low-level mercury detection instrument.
Features & Benefits
- Meets EPA & ATSDR Action Levels
- Battery Powered Sensor Regeneration
- 24 Hour Battery Life
- On-board Data Logging (20,000 data points) (communications option)
- Auto Time/Date Stamp
- USB data output (communications option)
- SCADA Interface Capabilities via 4-20ma
- Functional Test Kit (FTK) verifies instrument functions correctly between recommended annual factory calibrations
Quick specs
- Resolution: 0.01 µg/m3 (0.00001 mg/m3)
- Detection Range: 0.5 µg/m3 – 999 µg/m3 (0.0005 mg/m3 to 0.999 mg/m3)
- Data Storage Capacity: 20,000 samples
- Operating Environment: 0 – 45°C non-condensing, non-explosive
- Accuracy: 1.0 µg/m3, ±10%
25 µg/m3, ±5%
100 µg/m3, ±5%
- Typical Response Time: SURVEY MODE: 2 seconds
SAMPLE MODE: 16 seconds
- Flow Rate: 750 ± 50 cc/min (0.75 ± 0.05 liters/min)
- Power Requirements: 12 VDC for the instrument (provided by the internal battery, AC power
supply/charger, external battery pack or car accessory cable)
100-240VAC, 47-63Hz, 3.2A for the AC power supply/charger
- Internal Battery Pack: Rechargeable nickel metal hydride (NiMH)
- Dimensions: 11″ L x 6″ W x 6.5″ H
- Weight: 5.3 lbs.
- Output: Digital: USB serial data to PC, printer or USB flash drive
Analog: 4-20mA current loop (requires external power source)
accurate to 0.3% of output
- Warranty: 1 year, factory parts and labor